Patient monitor User’s manual
Chapter 2-4
The unit displaying this symbol contains an F-Type isolated
(floating) applied part providing a high degree of protection
against shock, and is defibrillator-proof.
Type CF applied part, defibrillation protected
The unit displaying this symbol contains an F-Type isolated
(floating) applied part providing a high degree of protection
against shock, and is defibrillator-proof.
The type CF applied parts provide a higher degree of protection
against electric shock than that provided by type BF applied
Attention! Please refer to the documents accompanying this
4. When a defibrillator is applied on a patient, the monitor may have transient
disorders in the display of waveforms. If the electrodes are used and placed
properly, the display of the monitor will be restored within 10s. During
defibrillation, please note to remove the electrode of chest lead and move
the electrode of limb lead to the side of the limb. The electrode of the
defibrillator should not come into direct contact with the monitoring
electrodes. Please ensure the monitor is reliably grounded and the
electrodes used repeatedly should be kept clean.
Warning: When conducting defibrillation, do not come into contact
with the patient, the bed and the monitor. Otherwise
serious injury or death could be resulted in.
5. To guarantee the safe operation of the monitor, the monitor is provided with
various replaceable parts, accessories and consuming materials (such as
sensors and their cables, electrode pads). Please use the products provided
or designated by the manufacturer.
6. Monitor only guarantees its safety and accuracy under the condition that it
is connected to the devices provided or designated by manufacturer. If the
monitor is connected to other undesignated electrical equipment or devices,
safety hazards may occur for causes such as the cumulating of the leakage