Patient monitor User’s manual
Chapter 5-5
alarm setting of parameter is independent from each other. Or you can select
【Main Menu】smartkey→【Alm Setup】to set the parameter alarm.
For the parameters whose alarm switch is set to ON, the alarm will be
triggered when at least one of them exceeds alarm limit. The following actions
1. Trigger audible and visual alarm according to alarm level;
2. Alarm message is displayed in the physiological alarm area;
3. The numeric of parameter in alarm flashes.
4. If alarm recording is on, the recorder starts alarm recording at set interval.
Select 【Alm Setup】→【Alm Limit Setup】 in each parameter setting
menu to set the alarm switch of them. You can select 【Alm Switch】to【On】
or【Off】. When a parameter alarm is off, a symbol “ ” will be displayed
near the parameter. If the alarms are turned off individually, they must be
turned on individually.
Select【Alm Setup】→【Alm Limit Setup】in each parameter setting menu
to set the alarm switch of them. You can select
【
Alm Level
】
in the setting
menu of each parameter, you can view and set the alarm levels of present
parameters. The level can be set to【Low】,【Med】or【High】.
Setting Individual Alarm Limits
The alarm limit can be set in the 【Alm Setup】menu of each parameter
setting menu, you can select【Hi Lmt】, 【Lo Lmt】to set the alarm limit for
each parameter. The alarm limit is adjustable, Physiological alarm will be
triggered when the measuring value exceeds the set limit.
Setting All Alarm Limits
Select【Main Menu】smartkey→【Alm Setup】→【Alm Limit Setup】
or 【 Alm 】 smartkey→【 Alm Limit Setup 】 to set the alarm limits of all