EasyManua.ls Logo

Arkray ADAMS A1C HA-8180V - Page 98

Arkray ADAMS A1C HA-8180V
246 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HA-8180V 2-41
7 Set the calibration information. [For the numeric button operation]
a On the standby screen, press .
b Set the HbA1C calibration conditions.
See the table below.
button: Moves the cursor to the next setup field.
Hyphen button: Changes the option.
c Press .
The current reagent information will appear.
d Below <Code>, check the current reagent code.
e To change the reagent code, enter the 10-digit code that is written
on the Standard value list.
If your entry is accepted, the lot number and expiration date will
be updated.
8 Start calibration.
a Start calibration.
For the barcode operation: Press .
For the numeric button operation: Press .
The message “Preparing for measurement” will appear and the rack will be transported to the aspiration position.
Cal
HbA1c calibration
10-06-2010 10:00
Standard value (mmol/mol)
L < 0> H < 0>
Conversion to NGSP: Y = a X + b
a: <0.0915>
b: <+2.15>
Calibration method
[Automatic ]
PrevItem BS OK Cancel
OK
Setup item Description
Standard value
(mmol/mol)
Enter the calibration standard values. Calibration standard values are written on the
Standard value list that comes with the calibrator.
L: 0 to 99 (Default: 0), H: 0 to 200 (Default: 0)
Conversion to NGSP Enter coefficients “a” and “b” of the conversion formula, “Y=aX+b”, to convert HbA1
C values
from the IFCC unit (mmol/mol) to the NGSP unit (%).
a: 0.0000 to 0.1500, b: -5.00 to 5.00 (Default: Those entered the last time)
Calibration method Select <Automatic>.
a: <0.0963>
b: +2.15
Calibration method
Automatic
Code Lot No. Expiry
<0123456789> 01234 2011-01
BS Start Cancel
NOTE:
If <Lot No.> and <Expiry> are not updated and still contain hyphens (---):
Your entry may be incorrect. Carefully check the reagent code and enter it again.
NOTE:
If an error message appears:
The reagent information is invalid. Press . Check the type, expiration date, manufacturing date of the
calibrator, and repeat from step 7-e.
Start
OK
2.8 HbA1C Calibration

Table of Contents

Related product manuals