Model 8640B TM 9-4935-601-14-7&P
Table 5-2. Post-Repair Test and Adjustments (1 of 4)
Assembly Repaired Performance Tests Adjustments
A1 Output Level Assy Harmonics Test (4-17) Check power supply voltages
Output Level Accuracy Test (5-25)
(4-23 or 4-24) Output Level Knob Adjust-
Output Level Flatness Test ment (5-30)
(4-25) Output Level Vernier and
Output Impedance Tests Meter Adjustment (5-31)
(4-26 and 4-27)
Output Leakage Test (4-29)
A2 Meter Switch/Detector Output Level Accuracy Test Check power supply voltages
Assy (upper 2 ranges only) (5-25)
A4 Meter/Annunciator Drive (4-24 or 4-25) Meter Adjustments (5-28)
Assy AM Sensitivity and Accuracy
Panel Meter M1 Test (meter only) (4-34)
FM Sensitivity and Accuracy
Test (meter only) (4-40)
A3 RF Oscillator Assy Frequency Range Test Check power supply voltages
(256-512 MHz only) (5-25)
(4-11)
V
T
Pot Adjustment (5-35)
Frequency Accuracy and Fine
V
T
Voltage Adjustment
Tune Test (fine tune only) (5-36)
(4-12)
Preliminary FM Adjustments
Frequency Stability Tests (if necessary) (5-38)
(4-13, 4-14, 4-15, and
4-16) FM Linearity Adjustment
(if necessary) (5-39) (5-44)
Harmonics Test (4-17)
FM Sensitivity Adjustment
Single Sideband Phase Noise (if necessary) (5-40)
Test (4-19) RF Oscillator End Stop
Residual FM Test (4-22) Adjustment (if necessary)
Output Level Flatness Test (5-42)
(256-512 MHz only) RF Oscillator Output Power
(4-25) Adjustment (if necessary)
Output Leakage Test (4-29) (5-43)
FM Distortion Test (4-39)
FM Sensitivity and Accuracy
Test (4-40)
Phase Lock Restabilization
Time Test (check only that
phase lock operates) (4-45)
5-4