EasyManua.ls Logo

artisan 8640B - Residual am Test; Residual am Test Setup

artisan 8640B
411 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 8640B TM 9-4935-601-14-7&P
PERFORMANCE TESTS
4-21. RESIDUAL AM TEST
SPECIFICATION:
Residual AM: (Averaged rms)
Post-detection Noise Bandwidth
300 Hz to 3 kHz 20 Hz to 15 kHz
>85 dB down >78 dB down
DESCRIPTION:
An rms voltmeter is calibrated with a measured amount of amplitude modulation from the Signal Generator. Then the
AM is removed and the generator's residual AM is read directly from the voltmeter.
Figure 4-8. Residual AM Test Setup
EQUIPMENT:
RMS Voltmeter.........................................................................................................HP 3400A
Detector....................................................................................................................HP 8471A
3 kHz Low-Pass Filter (LPF) ........................................................................CIR-Q-TEL 5 Pole
15 kHz Low-Pass Filter (LPF) ......................................................................CIR-Q-TEL 7 Pole
40 dB Amplifier...........................................................................................................HP 465A
Capacitor 5 µ F..................................................................................................HP 0180-2211
50 Ohm Load..........................................................................................................HP 11593A
PROCEDURE:
1. Connect equipment as shown in Figure 4-8 (with the generator connected to the rms voltmeter through the
detector, amplifier, 15 kHz LPF, and across the 50 ohm load). Set Signal Generator's controls as follows:
4-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents