Chapter 3 Parameters
Mini-Switch LB 471
Month/Day
Display or enter current date
The date is entered and displayed in the format MM.DD.
See code 01.
Hour/Minute
Display or enter current time
(Only in the professional mode)
The time is entered and displayed in the format hh.mm.
The time virtually has no influence on the decay compensation;
therefore, it does not really need to be set. However, it may be
used for test purposes.
Operation mode
Define operation mode
Input
Operation mode
0
Standard mode
1
Professional mode
See operation modes on page 17.
Detector code
Enter detector code for the detector used:
Input for
detector code
Designation
Detector type
0
LB 440X
NaI detector
(FSK)
0
LB 540X
23
LB 443X
Super Sens
(FSK)
23
LB 543X
98
SZ5 GHS 3171-2
GM detector
98
GHS 3172-2
99
SZ5 GHS 3171-1
99
GHS 3172-1
The correct detector code is required for the high voltage control
and the basic setting of the zero count rate. If the detector code is
changed, the standard calibration values of the selected detector
are automatically set in code 20, 21 and 22.
C o d e 0 2
C o d e 0 3
C o d e 0 4
C o d e 0 5