Chapter 3 Parameters
Mini-Switch LB 471
3.2 Standard Mode Code Table
Code
no.
Designation
Value range
Factory
setting
Page
00
Password
0000 - 9999
21
01
Year
1970 - 2099
Current year
21
02
Month / Day
01.01-12.31
Current date
21
04
Operation mode
0 = Standard
1 = Professional
0 - 1
0
22
05
Detector code
0 - 99
0 or 99
22
06
Nuclide
0=Co60, 1=Cs-137
0 - 1
0
22
10
Reading (%)
-999 - 9999
Reading
25
11
Reading in cps
averaged
0 - 999.9
Reading
25
16
Max. or min. limit value
switch
0=Max, 1=Min
0 1
0
27
20
Empty count rate (no input)
0 - 999.9
Reading
19
21
Full count rate (no input)
0 - 999.9
Reading
29
22
Zero count rate (no input)
0 - 9.999
Reading
30
32
Zero count rate
0 - 9.999
Depending on
detector code
32
39
Half-value layers
1-9
2
37
50
Limit switch software
1.00 - 9.99
Version
41
51
Detector software
(only FSK detector)
1.00 - 9.99
Reading
41
52
Detector temperature °C
(only FSK detector)
-40 80
Reading
41
53
Detector high voltage
(only FSK detector)
500 - 1300
-1
42
54
Detector HV-default
(only FSK detector)
500 - 1300
Set in factory
42
The same table is shown on page 75 with an empty column. In this
column you may enter your data after start-up.
Table 2:
Code table
for standard mode