Chapter 3 Parameters
Mini-Switch LB 471
Hysteresis (in %)
Hysteresis in percent
(Only in the professional mode)
The hysteresis increases the safety of the device by reducing the
risk of switching errors due to statistical fluctuations. The hystere-
sis lies symmetrically around the switching threshold. Data is en-
tered or displayed as percentage value between 0 and 100%.
If the hysteresis is set to automatic (factory setting), it is already
set up optimally and is constantly updated.
Perform a calibration in code 36 to check if your setting is correct.
Empty count rate (cps)
Display of empty count rate used by the device for
measurement
This parameter serves only for display.
The value is continuously decay-corrected.
Depending on the operation mode, this value is determined as
follows:
automatically following empty
calibration.
based on code 30 as soon as a
calibration in code 36 has been
performed.
Full count rate (cps)
Display of full count rate used by the device for
measurement
This parameter serves only for display.
The value is continuously decay-corrected.