Chapter 13 Technical Data
Mini-Switch LB 471
13.3 Shieldings
Approx. shielding thickness
(mm lead)
Approx. attenuation factor
for Cs-137
for Co-60
Dose rate on the surface of
the shielding (µSv/h)
with Cs-137 (A in MBq)
with Co-60 (A in MBq)
6.5 * 10
-6
* A
8.1 * 10
-3
* A
Dose rate in 1m distance
from the surface of the
shielding (µSv/h)
with Cs-137 (A in MBq)
with Co-60 (A in MBq)
1.4 * 10
-4
* A
1.1 * 10
-2
* A
5.4 * 10
-6
* A
1.7 * 10
-3
* A
1.1 * 10
-7
* A
1.5 * 10
-4
* A
Radius of controlled area, cm
with Cs-137 (A in MBq)
with Co-60 (A in MBq)
Dimensions (W x H x D), cm
with pneumatic locking
drive
18 * 20 * 20
18 * 20 * 39
24 * 27 * 27
24 * 27 * 46
30 * 39 * 38
30 * 39 * 57
13.4 Pneumatic Locking Drive
Signaling
OPEN/CLOSED
Option I
Option II
Option III
IP 65, 2 contact, 48 V DC, 1A
2 contact, max. 250 V AC, 1A, EEx e II T6
2 proximity switches, intrinsically safe feed
required