EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB4710-080 - 13.3 Shieldings

Default Icon
227 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 13 Technical Data
Mini-Switch LB 471
108
13.3 Shieldings
LB 7440 F
LB 7440 EF
LB 7442 F
LB 7442 EF
LB 7444
Approx. shielding thickness
(mm lead)
67
97
132
Approx. attenuation factor
for Cs-137
for Co-60
700
30
16000
180
650000
1800
Dose rate on the surface of
the shielding (µSv/h)
with Cs-137 (A in MBq)
with Co-60 (A in MBq)
1.6 * 10
-2
* A
1.6 * A
4.3 * 10
-4
* A
0.14 * A
6.5 * 10
-6
* A
8.1 * 10
-3
* A
Dose rate in 1m distance
from the surface of the
shielding (µSv/h)
with Cs-137 (A in MBq)
with Co-60 (A in MBq)
1.4 * 10
-4
* A
1.1 * 10
-2
* A
5.4 * 10
-6
* A
1.7 * 10
-3
* A
1.1 * 10
-7
* A
1.5 * 10
-4
* A
Radius of controlled area, cm
with Cs-137 (A in MBq)
with Co-60 (A in MBq)
0.74 * A1/2
6.5 * A1/2
0.15 * A1/2
2.6* A1/2
0.07 * A1/2
0.82* A1/2
Dimensions (W x H x D), cm
with pneumatic locking
drive
18 * 20 * 20
18 * 20 * 39
24 * 27 * 27
24 * 27 * 46
30 * 39 * 38
30 * 39 * 57
Operating temperature
max. 200 °C
max. 200 °C
max. 200 °C
Weight
31 kg
81 kg
170 kg
13.4 Pneumatic Locking Drive
Compressed air:
Min.
Max.
Connection
4 bar
7 bar
G 1/8
Air quality
clean, oil-free
Operating
temperature
-20 ... +80 °C
Signaling
OPEN/CLOSED
Option I
Option II
Option III
IP 65, 2 contact, 48 V DC, 1A
2 contact, max. 250 V AC, 1A, EEx e II T6
2 proximity switches, intrinsically safe feed
required

Table of Contents