Chapter 9 Maintenance
Mini-Switch LB 471
9.5.1 Replacing the Source
1
Open padlock
on the shielding.
Lever
Padlock
Safety screw
Knob
2
If necessary, unscrew safety screw
from lever.
3
Pull off knob
and rotate lever
to the right by 90 degrees
to the center position between “OPEN” and “CLOSED”. The Al-
len screw head of the source holder is now visible. (With some
versions the shutter plate has to be removed to be able to un-
screw the source holder.)
4
Using a socket wrench (size 12), unscrew the source holder
together with the source
.
Locking core
Radiation channel
Point sources
Source holder
Lever
Padlock
Cover plate
Figure 50:
Point source shielding
Figure 51:
Point source shielding -
sectional drawing