EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB4710-080 - Professional Mode Code Table

Default Icon
227 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Parameters
Mini-Switch LB 471
19
3.3 Professional Mode Code Table
Code
no.
Designation
Value
range
Factory
setting
Page
00
Password
0000 - 9999
21
01
Year
1970 - 2099
Current year
21
02
Month / Day
01.01-12.31
Current date
21
03
Hour / Minute
00.00-23.59
Current time
22
04
Operation mode
Standard/Professional
0 - 1
0
22
05
Detector code
0 - 99
99
22
06
Nuclide
0=Co60, 1=Cs-137
0 - 1
0
22
07
Automatic password
protection
0 - 9999
0
23
08
Warning relay as second
alarm relay
0 1
10 100
0
23
09
Alarm relay follows the error
relay
0 - 1
0
25
10
Reading (%)
-999 - 9999
Reading
25
11
Average count rate
0 - 999.9
Reading
25
12
Time constant (s)
0,1 999,9
-1
25
13
Live count rate
0 - 999.9
Reading
26
14
Maximum time constant (s)
0 - 999
999
26
15
Standard reading
10 - 11
10
26
16
Max. or min. limit value
switch
0=Max, 1=Min
0 - 1
0
27
17
Switching threshold (%)
0 - 100
-1
27
18
Switching threshold in (cps)
0 - 999.9
-1
28
19
Hysteresis (%)
0 - 999
-1
29
20
Empty count rate (no input)
0 - 999.9
Reading
19
21
Full count rate (no input)
0 - 999.9
Reading
29
22
Zero count rate (no input)
0 - 9.999
Reading
30
30
Empty count rate
0 - 999.9
20 GMZ
300 FSK
31
Full count rate
0 - 999.9
-1
31
32
Zero count rate
0 - 9.999
Depending on
detector code
32
33
Measuring path (in mm)
0 - 9999
0
33
34
Gas density (kg/m³)
0 - 9999
0
33
35
Bulk density (kg/m³)
0 - 9999
0
34
36
Compute
35.0135.08
Reading
35
37
Counting time for calibration (s)
5 - 600
60
36
38
Bulk cone diameter (mm)
0 - 9999
0
37
39
Half-value layers
1 - 9
2
37
Table 3:
Code table for
professional mode

Table of Contents