Chapter 3 Parameters
Mini-Switch LB 471
Signaling EVU excess temperature
Select relay to signal EVU excess temperature
(Only in the professional mode)
Signaling via error relay
Signaling via warning relay
The threshold for the excess temperature is defined in code 48.
The warning relay cannot be used for signaling if code 8 is set to
1.
EVU limit temperature
Enter EVU limit temperature (in °C)
(Only in the professional mode)
Switching threshold for signaling excess temperature in the evalu-
ation unit electronics. The signaling can be used to ensure the reli-
ability of the measurement. The signal is indicated on the display
and, if necessary, also by a relay. The relay for signaling is select-
ed in code 47.
Limit switch software version
Show software version of evaluation unit
A new software version can be uploaded only by means of a pro-
gramming device or by BERTHOLD TECHNOLOGIES in Bad Wildbad.
Detector software version
Show software version of FSK detector (only FSK detectors)
If necessary, the software in the detector can be replaced by an
experienced user.
Detector temperature
Show current temperature inside the detector
(only FSK detectors)
The detector electronics includes a temperature sensor. An alarm
is triggered when reaching the temperature set in code 46. The
maximum permissible ambient temperature is listed in the tech-
nical data in the hardware manual.