Chapter 4 Getting Started
Mini-Switch LB 471
55
5
Code 16: Select min./max. limit value switch
Maximum value limit switch
Input
Alarm relay function
0
Maximum value switch
1
Minimum value switch
6
Code 32: Read in zero count rate
Example:
50 cps as natural environmental radiation
Input
Function
„Cal“ but-
ton
Read in count rate
7
Code 39: Enter half-value layers
Example: 2 half-value layers
(sufficient in most cases)
Input
Function
1-9
Number of half-value layers
This completes the basic setting.
The measurement can now be calibrated.
C 39
2
C 32
50
C 16
0