Data Analysis Windows
94
Well spreadsheet. Shows a spreadsheet listing the allelic discrimination data collected
in the selected wells
Figure 94. Layout of the Allelic Discrimination tab in the Data Analysis window.
Adjusting Data for Allelic Discrimination
The software automatically assigns a genotype to wells with unknown samples based on the
positions of the vertical and horizontal threshold bars, and then lists genotype calls in the
spreadsheet view. To automatically call genotypes, the software uses positive controls (when
available), or estimates the thresholds. The software takes an average C(t) or RFU for the
positive controls to automatically set the threshold lines for discriminating the alleles.
Adjust the position of the threshold bars by clicking and dragging them, and the software
automatically adjusts the calculations to make new genotype assignments:
If the experiment contains three controls in the plate, then the position of the
threshold bars is based on the mean and standard deviation of the RFU or C(t) of the
controls
If the number of controls is less than three, then the position of the threshold bars is
determined by the range of RFU or threshold cycle values in the selected fluorophore
Adjust allelic discrimination data by following any of these methods:
Click and drag the threshold bars in the Allelic Discrimination chart to adjust the calls
in the spreadsheet
Select a fluorophore for each axis in the chart (X: and Y:) in the settings options on
the bottom right of the window

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BIO RAD CFX96 and is the answer not in the manual?

BIO RAD CFX96 Specifications

General IconGeneral
Temperature Uniformity±0.4 °C
SoftwareCFX Maestro Software
Detection MethodReal-time PCR (qPCR)
Sample Capacity96-well plate
Excitation SourceLED
Ramp RateUp to 5°C/sec
Heating Rate5°C/sec
SensitivitySingle copy detection
Power Requirements100-240 V, 50/60 Hz
Temperature Range0–100 °C

Related product manuals