3. Protection and Error Messages
3.1 Overvoltage Protection (OVP)
When OVP is activated and the voltage measured exceeds the set point, the
system will enter the “Overvoltage Protection” mode that will shut off the
output and show OVP on the display. Press ESC to reset the protection mode
and deactivate the the buzzer.
3.2 Overcurrent Protection (OCP)
When OCP is activated and the current measured exceeds the set point of
protected current, system will enter the Overcurrent Protect” mode that will
shut off the output and show OCP on the display. Press ESC to reset the
protection mode and deactivate the buzzer.
3.3 Overpower Protection (OPP)
When OPP is activated and the power measured exceeds the set point, the
system will enter the “Overpower Protection” mode that will shut off the output
and display OPP on the screen. Press ESC to reset the protection mode and
deactivate buzzer.
= .
.
= .
.