58
characters at the end of each command when programming.
6.3 The Error/Event List
SCPI interface can offer an error/event list that contains up to 10 errors/events.
Users can read the errors/events through the āerror?ā command in a first-in
first-out manner. Once an error/event is read, the read process will clear it from
the memory. To clear all errors/events from the memory, the ā*CLSā command
is used.
-000 No error
-001 Command error
-002 Execution error
-003 Query error
-004 Input Range error
-005 Parallel/Series function, Error mode
-006 Parallel/Series function, Multi-Master
-007 Parallel/Series function, No Slave found
-008 Parallel/Series function, Communication with Slave A
error
-009
Parallel/Series function, C
ommunication with Slave B
error
-010 Parallel/Series function, C
ommunication with Slave C
error
-011 Parallel/Series function, Sync. signal error when
output on
-012 Parallel/Series function, Sync. signal error when
output off