5.13 Error Codes 5.13 Error Codes
CapacityTester
error code
in DiagnosticTool
error code
in display
Error description, repair
0xAA1001
0xAA3001
0xAA3002
0xAA4000
0xAA5000
0xAC0105
0xAC0106
0xAC0107
0xAC0108
0xAC0109
0xAC0110
–
Internal software error. Print the diagnostic report, clear the error, and
restart the device. Perform software update.
0xAD0000 –
Incorrect hardware version. Check the software version of the Capacity-
Tester. Print the diagnostic report, clear the error, and restart the device.
Perform software update.
0xAE0000
0xAE1000
–
Internal communication error. Print the diagnostic report, clear the error,
and restart the device. Perform software update.
0xAE1001 –
Incorrect software variant. Print the diagnostic report, clear the error, and
restart the device. Perform software update.
0xAE1002
0xAE1003
–
Incorrect hardware variant. Print the diagnostic report, clear the error, and
restart the device. Perform software update.
0xAE2000 –
Incorrect battery item number. Print the diagnostic report, clear the error,
and restart the device. Perform software update.
PC_TOOL FAULT
No. 0
– No error in the error memory.
PC_TOOL FAULT
No. 1
–
CAN communication with the connected battery is not possible. Check the
connection between the battery and the CapacityTester, if the battery is
connected correctly, check the battery fault.
PC_TOOL FAULT
No. 2
–
The selected battery type does not match the connected battery.
Adjust the battery type in DiagnosticTool.
PC_TOOL FAULT
No. 3
–
CAN communication was interrupted during the measurement. Check the
connection between the battery and the CapacityTester. If the connection
is correct, check the error memory of the battery.
PC_TOOL FAULT
No. 4
–
The charge level of the battery is too low. The battery must be charged
before starting the measurement.
PC_TOOL FAULT
No. 5
– Cooling the CapacityTester does not work properly, restart the system.
5
SERVICE