6 Detecting and correcting faults
104 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
Heading
Fault message
ARC PLUS Fault
Check if the argon bottles are connected and open. Empty
bottles should be replaced.
You can connect two argon bottles. A change to the
replacement bottle occurs automatically.
ARC PLUS
Warning
The filling level of the argon bottle is low.
Please make sure that a replacement is available.
You can connect two argon bottles. The unit shifts
automatically to the second bottle.
ARC PLUS Fault
The argon bottle is empty.
Connect a replacement bottle to enable activation.
You can connect two argon bottles. The unit shifts
automatically to the second bottle
ARC PLUS Notice
The argon bottle is empty. The unit has shifted automatically
to the replacement bottle.
Please make sure that a replacement is available.
ARC PLUS Notice
The argon bottle is empty. The unit shifts automatically to the
replacement bottle.
Please make sure that a replacement is available.
ARC PLUS Fault
Please check if the argon bottles are connected and open.
Empty bottles should be replaced.
Subsequently restart ARC PLUS by activating the flashing
"Purge" button.
TSI ARC PLUS
Warning
The annual Technical Safety Inspection (TSI) for ARC PLUS
is due.
Plug'n Cut
COMFORT Notice
The lifetime of the instrument is ending soon.
Please order a replacement in good time.
Any use of the instrument beyond its lifetime is not covered
by warranty. Please contact your BOWA dealer in good time
to purchase a new instrument.
Plug'n Cut
COMFORT
Warning
The maximum lifetime of the instrument has been reached.
Any further use is not covered by warranty. The maximum
service lifetime of the instruments must not be exceeded, in
order to guarantee safe usage. Any further use is at the
user's risk.
Plug'n Cut
COMFORT
Warning
A software update is necessary to use Plug'n Cut COMFORT
with this instrument. Only carry out manual settings at this
instrument. Please contact the Technical Support. Contact:
MENU - SERVICE.
Plug'n Cut
COMFORT
Warning
Unable to load the preference parameters of the COMFORT
instrument.
Configure the instrument settings manually.
Please contact the Technical Support.
Contact: MENU - SERVICE.