EasyManua.ls Logo

Bowa ARC 400 - Page 117

Bowa ARC 400
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Technical specifications
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 117
Mode icon
Description
CCS
ARC CONTROL
Form of HF voltage
Max. power
output
Peak voltage
Default
values
Rated load resistance
Modulation
Effect
Power range
Effect
Max. Watt
Frequency
Duty cycle
Modes Monopolar Coagulation
Moderate
sinusoidal
constant
1
2
3
1 W
120 W
250 Vp
2
60
75 Ω
---
---
Forced non cutting
pulsed
modulated
-
1 W - 80 W
3.5 kVp
---
50
1000 Ω
20 kHz
1 pulse
Forced mixed
sinusoidal
modulated
1
2
3
1 W
120 W
1.5 kVp
2.0 kVp
2.5 kVp
2
60
500 Ω
30 kHz
sinusoidal
1 pulse
Forced cutting
sinusoidal
modulated
1
2
3
4
1 W
250 W
1.5 kVp
1.5 kVp
1.3 kVp
1.3 kVp
2
80
500 Ω
20 kHz
30 %
35 %
40 %
50 %
Spray
pulsed
modulated
1
2
3
4
1 W
120 W
3.0 kVp
3.8 kVp
4.6 kVp
5.0 kVp
2
80
500 Ω
20 kHz
sinusoidal 1
pulse
1 Impuls
1 Impuls
1 Impuls
Argon
pulsed
modulated
-
1 W
120 W
4.6 kVp
---
80
500 Ω
20 kHz
1 pulse
Argon flexible
pulsed
modulated
-
1 W
120 W
4.4 kVp
---
20
500 Ω
Power
dependent
1 kHz - 20
kHz
1 pulse
Argon flex. pulse
pulsed
modulated
1
2
3
1 W 80 W
4.4 kVp
2
20
500 Ω
Power
dependent
1 kHz - 20
kHz
1 pulse
1 pulse
1 pulse
Resection
sinusoidal
modulated
-
1 W -120 W
2.2 kVp
---
60
500 Ω
30 kHz
sinusoidal 1
pulse

Table of Contents

Related product manuals