10 Technical specifications
118 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
Mode icon
Description
CCS
ARC CONTROL
Form of HF voltage
Max. power
output
Peak voltage
Default
values
Rated load resistance
Modulation
Effect
Power range
Effect
Max. Watt
Frequency
Duty cycle
Modes Monopolar Coagulation
Cardiac Mammary
sinusoidal
modulated
-
1 W - 60 W
1.8 kVp
---
15
500 Ω
30 kHz
sinusoidal 1
pulse
Cardiac Thorax
sinusoidal
modulated
-
1 W
100 W
1.8 kVp
---
40
500 Ω
30 kHz
sinusoidal 1
pulse
SimCoag
sinusoidal
modulated
pulsed
modulated
pulsed
modulated
1
2
3
1 W
120 W
2.0 kVp
2.5 kVp
4.6 kVp
2
60
500 Ω
30 kHz
30 kHz
20 kHz
sinusoidal 1
pulse
sinusoidal 1
pulse
1 pulse
Gastro Coag
sinusoidal
modulated
1
2
3
1 W - 50 W
1.8 kVp
2.2 kVp
2.8 kVp
2
15
500 Ω
30 kHz
sinusoidal
1 pulse
Laparoscopy
sinusoidal
modulated
-
1 W
120 W
1.8 kVp
---
60
500 Ω
20 kHz
5%
Modes Bipolar Cutting
Standard
Yes
Yes
sinusoidal
constant
-
1 W
200 W
400 Vp
---
100
75 Ω
---
---
Bipolar resection
Yes
Yes
sinusoidal
constant
1
2
3
250 W
500 Vp
2
---
75 Ω
---
---
Initial incision phase
860 W
Bipolar scissors
sinusoidal
constant
-
1 W
120 W
200 Vp
---
40
75 Ω
---
---
Vaporisation
Yes
Yes
sinusoidal
constant
1
2
3
300 W
300 W
400 W
350 Vp
400 Vp
450 Vp
2
---
75 Ω
---
---