10 Technical specifications
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 119
Mode icon
Description
CCS
ARC CONTROL
Form of HF voltage
Max. power
output
Peak voltage
Default
values
Rated load resistance
Modulation
Effect
Power range
Effect
Max. Watt
Frequency
Duty cycle
Modes Bipolar Coagulation
Standard
forceps
sinusoidal
constant
-
1 W
120 W
150 Vp
---
40
50 Ω
---
---
Standard
forceps AUTO
sinusoidal
constant
-
5 W
120 W
150 Vp
---
40
50 Ω
---
---
Micro forceps
sinusoidal
constant
-
0.1 W
40 W
90 Vp
---
10
50 Ω
---
---
Forceps forced
sinusoidal
modulated
-
1 W
100 W
550 Vp
---
50
50 Ω
20 kHz
10%
LIGATION
sinusoidal
modulated
-
200 W
190 Vp
---
---
25 Ω
1 - 2 Hz
sinusoidal
TissueSeal PLUS
sinusoidal
modulated
-
200 W
190 Vp
---
---
25 Ω
1 - 2 Hz
sinusoidal
Bipolar scissors
sinusoidal
constant
-
1 W
120 W
200 Vp
---
40
75 Ω
---
---

Table of Contents