Loading...
10 Technical specifications
120 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
Mode icon
Description
CCS
ARC CONTROL
Form of HF voltage
Max. power
output
Peak voltage
Default
values
Rated load resistance
Modulation
Effect
Power range
Effect
Max. Watt
Frequency
Duty cycle
Modes Bipolar Coagulation
Laparoscopy
sinusoidal
constant
-
1 W
120 W
150 Vp
---
50
50 Ω
---
---
Laparoscopy Micro
sinusoidal
constant
-
1 W
100 W
110 Vp
---
40
25 Ω
---
---
Bipolar resection
sinusoidal
constant
1
2
3
4
125 W
200 W
275 W
350 W
190 Vp
2
---
25 Ω
---
---
SimCoag
sinusoidal
modulated
-
5 W - 60 W
550 Vp
---
40
50 Ω
20 kHz
50%
Vaporisation
sinusoidal
constant
sinusoidal
modulated
sinusoidal
modulated
1
2
3
250 W
190 Vp
400 Vp
500 Vp
2
---
25 Ω
-
20 kHz
20 kHz
----
50%
50%
The max. values are not necessarily created at rated load resistance.
The HF power is subject to a tolerance limit of ± 20%.