10 Technical specifications
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 127
Monopolar Cutting Laparoscopy
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting
Laparoscopy"
= 100 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting
Laparoscopy"
= 200 W
Diagram of power output P [W] as a
function of the setting "Monopolar
Cutting Laparoscopy"
Rated load resistance= 500
Effect
U (Vp)
1
400
2
450
3
560
4
650
5
650
6
700
7
700
8
700
9
750
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Cutting Laparoscopy" (idle mode)