10 Technical specifications
128 Operating Manual ARC 400 BOWA-IFU-11779-ARC400-V2.1.0-S0-EN
Monopolar Cutting GastroLOOP 1, 2, 3
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting
GastroLOOP 1, 2, 3"
coag phase
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Monopolar Cutting
GastroLOOP 1, 2, 3"
cut phase
Effect
P (W)
coag phase
P (W)
cut phase
1
17
400
2
21
400
3
26
400
4
30
400
5
35
400
Table of power output P [W] as a
function of the setting "Monopolar
Cutting GastroLOOP 1, 2, 3"
Rated load resistance= 500
Effect
U (Vp)
1
750
2
750
3
750
4
750
5
750
Table of HF output voltage U [Vp] as a
function of the setting "Monopolar
Cutting GastroLOOP 1, 2, 3"
(idle mode)