10 Technical specifications
BOWA-IFU-11779-ARC400-V2.1.0-S0-EN Operating Manual ARC 400 151
Bipolar Coagulation Forceps forced
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Bipolar Coagulation forceps
forced"
= 50 W
Measurement at ohmic resistances
Diagram of power output P [W] as a
function of the load resistance R [] for
the setting "Bipolar Coagulation forceps
forced"
= 100 W
Diagram of power output P [W] as a
function of the setting " Bipolar
Coagulation forceps forced "
Rated load resistance = 50
HF output voltage U [Vp] for the setting
" Bipolar Coagulation forceps forced "
(idle mode)
= 550 Vp