EasyManuals Logo

Bruker HPPR/2 User Manual

Bruker HPPR/2
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
MRI Active Detuning Module
60 / 92 H172204_1_001
10.2 Description
Figure10.2: Block Diagram: MRI ADM-CRP Module (ECL > 03)
10.2.1 Input Pulses
The ADM module uses seven input pulses used for PIN driver, tune/match driver and RGP
output pulse control:
RGP_HPPR: common RGP for all preamplifier modules.
EXT_RGP(1:2)~: discrete RGP from AV4 TRX.
AD-CTRL(1:4)~
2
: discrete transmitter blanking pulse from AV4 AD-CTRL Board.
2
Formerly BLNKTR(1:4)~

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker HPPR/2 and is the answer not in the manual?

Bruker HPPR/2 Specifications

General IconGeneral
BrandBruker
ModelHPPR/2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals