MRI Active Detuning Module
60 / 92 H172204_1_001
10.2 Description
Figure10.2: Block Diagram: MRI ADM-CRP Module (ECL > 03)
10.2.1 Input Pulses
The ADM module uses seven input pulses used for PIN driver, tune/match driver and RGP
output pulse control:
• RGP_HPPR: common RGP for all preamplifier modules.
• EXT_RGP(1:2)~: discrete RGP from AV4 TRX.
• AD-CTRL(1:4)~
2
: discrete transmitter blanking pulse from AV4 AD-CTRL Board.
2
Formerly BLNKTR(1:4)~