Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
8-12
MEASure[:SCALar]:RLC:SET:LS?
FETCh[:SCALar]:RLC:SET:LS?
Description: Return the Ls measurement value of RLC CP mode
Query Syntax : MEASure[:SCALar]:RLC:SET:LS?
FETCh[:SCALar]:RLC:SET:LS?
Response Parameter : <NR2>
Example : MEASure[:SCALar]:RLC:SET:LS?
MEASure[:SCALar]:RLC:SET:RL?
FETCh[:SCALar]:RLC:SET:RL?
Description: Return the R
L
measurement value of RLC CP mode
Query Syntax : MEASure[:SCALar]:RLC:SET:RL?
FETCh[:SCALar]:RLC:SET:RL?
Response Parameter : <NR2>
Example : MEASure[:SCALar]:RLC:SET:RL?
MEASure[:SCALar]:RLC:SET:RS?
FETCh[:SCALar]:RLC:SET:RS?
Description: Return the Rs measurement value of RLC CP mode
Query Syntax : MEASure[:SCALar]:RLC:SET:RS?
FETCh[:SCALar]:RLC:SET:RS?
Response Parameter : <NR2>
Example : MEASure[:SCALar]:RLC:SET:RS?
8.6.3 LOAD Subsystem
LOAD
[:STATus] Set load ON / OFF
[LOAD:]
ABA Set ABA mode
MODE Set CC / CP / CV / CR mode
SCIRcuit Set short ON / OFF
PROTect Queue protection status
CLEar Reset protection status
LOAD[:STATus]
Description: The LOAD command makes the electronic load active/on or inactive/off.
Setting Syntax : LOAD[:STATus]?
Setting Parameter : 0 | OFF, 1 | ON
Query Syntax : LOAD[:STATus]?
Response Parameter: 0 | 1
Example : LOAD ON
[LOAD:]ABA
Description: ABA Mode setting. If the setting value is set to 'OFF', then the system will not
automatically adjust the bandwidth. If the setting value is set to 'ON' the system will
automatically adjust the bandwidth upon connecting the UUT to the AC Load. Before drawing
a current, the system will calculate the inductance of the UUT to adjust the bandwidth
accordingly before loading.
Setting Syntax : [LOAD:]ABA?
Setting Parameter : 0 | OFF, 1 | ON