Verification
4. Please confirm if the values of Meter, Load & DVM meet to that in Table 10-7.
Table 10-7
Source Setting
V_Mon
V Specification Specification
45/440
10.5 DC Performance Test
10.5.1 I monitor Measurement Verification
DC Source wiring as Figure 10-2, switch AC Load to DC mode, the procedures are as follows.
1. Press CONF./LOCAL.
2. Press 1 and ENTER to enter SETUP.
3. Move the cursors of ▲ and ▼ to “1-1. MODE SELECT”.
4. Press direction keys of ◄, ► or turn rotary knob to select “DC” and then press ENTER.