6. AC Source settings are shown as Table 10-5. Please follow the model to set W, PF &
I
p(max). According to the model to confirm if the values of Meter & Load meet to that in
Table 10-5.
Table 10-5
10.4.5 THD Measurement Verification
AC Source sets output voltage waveform to sine, AC Load enters CC mode and no loading.
Select ▼ at the upper side of rotary knob to press for twice, observe if THD measurement
value of Load and Meter panel meet to Table 10-6.
I
=
0
. 0
0
T
H
D
=
0
. 0
0
I
=
0
. 0
0
Table 10-6
Model
THD THD(%) Specification
150
AC Load
10.4.6 V monitor, Voltage Measurement Verification
1. Press CC to enter CC mode.
2. Set AC SOURCE to series file to output current limit 1A, other settings are as Table 10-7.
3. Switch DVM measurement terminal to Front and DVM range to AC for measuring V_Mon
voltage.
Model
W PF Ip(max)
50 45/440
54
63802
108
63803
135
63804
4500 1
0.99~1.01