Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
10-6
6. AC Source settings are shown as Table 10-5. Please follow the model to set W, PF &
I
p(max). According to the model to confirm if the values of Meter & Load meet to that in
Table 10-5.
Table 10-5
10.4.5 THD Measurement Verification
AC Source sets output voltage waveform to sine, AC Load enters CC mode and no loading.
Select at the upper side of rotary knob to press for twice, observe if THD measurement
value of Load and Meter panel meet to Table 10-6.
I
s
e
t
=
0
.
5
0
0
C
F
=
1
.
4
1
4
P
F
=
1
.
0
0
0
I
p
(
m
a
x
)
=
1
0
.
0
U
N
I
T
:
A
I
P+
=
0
. 0
0
T
H
D
=
0
. 0
0
I
P-
=
0
. 0
0
A
C
P
R
I
O
R
I
T
Y
:
C
F
Table 10-6
Source Setting
Meter & Load Measurement
Model
Voltage
(V)
Frequency
(Hz)
THD THD(%) Specification
150
45
10%
9.9~10.1
AC Load
440
40%
39.6~40.4
10.4.6 V monitor, Voltage Measurement Verification
1. Press CC to enter CC mode.
2. Set AC SOURCE to series file to output current limit 1A, other settings are as Table 10-7.
3. Switch DVM measurement terminal to Front and DVM range to AC for measuring V_Mon
voltage.
Source Setting
Load Setting
Meter & Load Measurement
Model
Voltage
(V)
Frequency
(Hz)
W PF Ip(max)
P
Specification
PF
Specification
50 45/440
50
0.2
54
44.5~55.5
0.19~0.21
63802
1800
0.6
1791~1809
0.59~0.61
50
0.2
108
39.1~60.9
0.19~0.21
63803
3600
0.8
3582~3618
0.79~0.81
50
0.2
135
36.4~63.3
0.19~0.21
63804
4500 1
4477.5~4522
.5
0.99~1.01