Remote Operation
Description: It sets the value C, this command is only operating in AC Inrush Current Mode.
Setting Syntax : [LOAD:] INRush:CAP
Setting Parameter : <NR1>, Range: 100 ~ 9999, Unit: uF
Query Syntax : [LOAD:] INRush:CAP?
Response Parameter: <NR1>
Example : INR:CAP 470
[LOAD:] INRush:LS
Description: It sets the value LS, this command is only operating in AC Inrush Current Mode.
Setting Syntax : [LOAD:] INRush:LS
Setting Parameter : <NR1>, Range: 0 ~ 9999, Unit: uH
Query Syntax : [LOAD:] INRush:LS?
Response Parameter: <NR1>
Example : INR:LS00
[LOAD:]INRush:RL
Description: It sets the value RL, this command is only operating in AC Inrush Current Mode.
Setting Syntax : [LOAD:]INRush:RL
Setting Parameter : <NR2>, 63804 Range: 1.11 ~ 9999.99, 63803 Range: 1.39 ~ 9999.99,
63802 Range: 2.77 ~ 9999.99, Unit: Ohm
Query Syntax : [LOAD:]INRush:RL?
Response Parameter: <NR2>
Example : INR:RL 50
[LOAD:]INRush:RS
Description: It sets the value RS, this command is only operating in AC Inrush Current Mode.
Setting Syntax : [LOAD:]INRush:RS
Setting Parameter : <NR2>, Range: 0 ~ 9.999, Unit: Ohm
Query Syntax : [LOAD:]INRush:RS?
Response Parameter: <NR2>
Example : INR:RS 0.1
[LOAD:] INRush:PHASe
Description: It sets the value phase, this command is only operating in AC Inrush Current
Mode.
Setting Syntax : [LOAD:]INRush:PHASe
Setting Parameter : <NR2>, Range: 0 ~ 359, Unit: Degree
Query Syntax : [LOAD:]INRush:PHASe?
Response Parameter: <NR2>
Example : INR:PHAS 60
8.6.3.9 DC Rectified Subsystem
[LOAD:]
SYNC
:DC Select line frequency or set frequency mode
FREQ (DC Rectified mode only)
[:LEVel]
[:AMPLitude]
:DC Set the frequency
[LOAD:]SYNC:DC