Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
[LOAD:] RLC:LS
Description: It sets the value Ls, this command is only operating in AC RLC Mode.
Setting Syntax : [LOAD:] RLC:LS
Setting Parameter : <NR1>, Range: 0 ~ 9999, Unit: uH
Query Syntax : [LOAD:] RLC:LS?
Response Parameter: <NR1>
Example : RLC:LS 200
[LOAD:] RLC:RL
Description: It sets the value RL, this command is only operating in AC RLC Mode.
Setting Syntax : [LOAD:] RLC:RL
Setting Parameter : <NR2>, 63804 Range: 1.11 ~ 9999.99, 63803 Range: 1.39 ~ 9999.99,
63802 Range: 2.77 ~ 9999.99, Unit: Ohm
Query Syntax : [LOAD:] RLC:RL?
Response Parameter: <NR2>
Example : RLC:RL 50
[LOAD:] RLC:RS
Description: It sets the value RS, this command is only operating in AC RLC Mode.
Setting Syntax : [LOAD:] RLC:RS
Setting Parameter : <NR2>, Range: 0 ~ 9.999, Unit: Ohm
Query Syntax : [LOAD:] RLC:RS?
Response Parameter: <NR2>
Example : RLC:RS 0.1
[LOAD:] RLC:POW
Description: It sets the value C, this command is only operating in AC RLC CP Mode.
Setting Syntax : [LOAD:] RLC:POW
Setting Parameter : <NR2>, Range: 100 ~ 4500, Unit: W
Query Syntax : [LOAD:] RLC:POW?
Response Parameter: <NR2>
Example : RLC:POW 1000
[LOAD:] RLC:PFAC
Description: It sets the value PF, this command is only operating in AC RLC CP Mode.
Setting Syntax : [LOAD:] RLC:PFAC
Setting Parameter : <NR2>, Range: 0.400 ~ 0.750(AC RLC CP Mode)
Query Syntax : [LOAD:] RLC:PFAC?
Response Parameter: <NR2>
Example : RLC:PFAC 0.6
8.6.3.8 Inrush Current Subsystem
[LOAD:]
INRush
:CAP Set the C (AC RLC Inrush Current Mode)
:LS Set the L
S
(AC RLC Inrush Current Mode)
:RL Set the R
L
(AC RLC Inrush Current Mode)
:RS Set the R
S
(AC RLC Inrush Current Mode)
:PHASe Set the start phase (AC RLC Inrush Current Mode)
[LOAD:] INRush:CAP