Local Operation
3.17.4 CV Mode
Under timing mode to enter CV timing mode press the CV key.
In this mode the loading voltage (V
set
), maximum current value I
p
(max) , as well as the timeout
time and cutoff voltage of the UUT can be set.
V
= 0
. 0
0
T
=
0
0
: 0 0 : 0
0 : 0
0
0
m
s
The procedures of setting V
set
= 200.00, I
p
(max) = 29.0, Timeout = 00:00:05, and Cutoff V =
8.5, are described as below:
1. Move the cursor to the command line of “V
set
=”.
2. Press 2 , 0 , 0 then press ENTER to change the value to “200.00”.
3. Move the cursor by pressing ( ◄, ► ) to the command line of “I
p
(max)=”.
4. Press 2 , 9 then press ENTER to change the value to “29.0”.
5. Move the cursor by pressing ( ◄, ► ) to the command line of “TIMEOUT=00:00:00”.