EasyManua.ls Logo

Chroma 63800 Series - Voltage Subsystem; Crest Factor Subsystem

Chroma 63800 Series
198 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
8-18
~ 2500.00
DC mode, 63804: 1.00~2500.00 , 63803: 1.25~2500.00, 63802:
2.22~2500.00
Example : RES:LIM?
8.6.3.4 Voltage Subsystem
[LOAD:]
VOLTage
[:LEVel]
[:AMPLitude] Set the RMS voltage
:DC
[LOAD:]VOLTage[:LEV][:AMPL]:DC
Description: This command sets the DC voltage of the electronic load. The electronic load
insures that the load voltage stays constant at the set value.
Setting Syntax : [LOAD:]VOLTage[:LEV][:AMPL]:DC
Setting Parameter : <NR2>, Range: 7.50 ~ 500.00, Unit: V
Query Syntax : [LOAD:]VOLTage[:LEV][:AMPL]:DC?
Response Parameter: <NR2>
Example : VOLT:DC 20
8.6.3.5 Crest Factor Subsystem
[LOAD:]
CFAC
[:LEVel]
[:AMPLitude]
[:AC] Set the Crest Factor (AC CC/CP Mode)
:DC Set the Crest Factor (DC Rectified Mode)
:LIMit? Return the setting range of CF
[LOAD:]CFAC[:LEV][:AMPL][:AC]
Description: This command sets the crest factor of the electronic load. The electronic load will
attempt to program the crest factor to a value as close as possible to the set value. The ability
to do so will be affected by the mode and the power factor setting.
Setting Syntax : [LOAD:]CFAC[:LEV][:AMPL][:AC]
Setting Parameter : <NR2>, Range: 1.414 ~ 5.000
Query Syntax : [LOAD:]CFAC[:LEV][:AMPL][:AC]?
Response Parameter: <NR2>
Example : CFAC 1.414
[LOAD:]CFAC[:LEV][:AMPL]:DC
Description: It sets the value CF, this command is only operating in DC Rectified Mode.
Setting Syntax : [LOAD:]CFAC[:LEV][:AMPL]:DC
Setting Parameter : <NR2>, Range: 1.414 ~ 5.000
Query Syntax : [LOAD:]CFAC[:LEV][:AMPL]:DC?
Response Parameter: <NR2>
Example : CFAC:DC 1.414

Table of Contents

Related product manuals