Verification
Table 10-10
Source Setting Load Setting
Model
R (Ω) Ip(max) R Specification (Ω)
18
54
63802
36
108
63803
50
135
63804
10.5.4 CP Mode Operation Verification
1. Press CP to return to main screen of DC CP mode.
2. High current DC Source settings are shown as Table 10-11. Please follow the model
under CP mode to set the loading watt & I
p(max). According to the model to confirm if the
values of Meter & Load meet to that in Table 10-11.
Source Setting Load Setting
Model
W Ip(max)
100 50
54
63802
108
63803
135
63804
Table 10-11
10.5.5 CV Mode Operation Verification
1. Press CV to return to main screen of DC CV mode.
2. High current DC Source settings are shown as Table 10-12. Set loading voltage & I
max
under CV mode and confirm if the values of Meter & Load meet to that in Table 10-12.
Table 10-12
Source Setting Load Setting
Current (A) V Imax V Specification