Verification
10-9
Table 10-10
Source Setting Load Setting
Meter & Load
Measurement
Model
Voltage
(V)
Current
(A)
R (Ω) Ip(max) R Specification (Ω)
100
18
100
54
81.286~129.908
63802
50
3
2.965~3.036
100
36
100
108
71.174~168.067
63803
50
1.5
1.484~1.517
100
50
100
135
66.445~202.020
63804
50
1.2
1.187~1.213
10.5.4 CP Mode Operation Verification
1. Press CP to return to main screen of DC CP mode.
2. High current DC Source settings are shown as Table 10-11. Please follow the model
under CP mode to set the loading watt & I
p(max). According to the model to confirm if the
values of Meter & Load meet to that in Table 10-11.
Source Setting Load Setting
Meter & Load
Measurement
Model
Voltage
(V)
Current
(A)
W Ip(max)
P Specification
(W)
100 50
50
54
44.5~55.5
63802
1800
1791~1809
50
108
39.1~60.9
63803
3600
3582~3618
50
135
36.4~63.3
63804
4500
4477.5~4522.5
Table 10-11
10.5.5 CV Mode Operation Verification
1. Press CV to return to main screen of DC CV mode.
2. High current DC Source settings are shown as Table 10-12. Set loading voltage & I
max
under CV mode and confirm if the values of Meter & Load meet to that in Table 10-12.
Table 10-12
Source Setting Load Setting
Meter & Load
Measurement
Voltage
(V)
Current (A) V Imax V Specification
10
1
7.5
2
7.09~7.41
80
14
70
15
69.54~70.46