Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
10-8
5. Press CC to return to main screen of DC mode.
6. High current DC Source settings are shown as Table 10-8, switch DVM2 measurement
terminal to Rear. Please follow the model under CC mode to set loading current & I
p(max).
According to the model to confirm if the values of DVM meet to that in Table 10-8.
Table 10-8
Source Setting
Load Setting
DVM
Model
Voltage
(V)
Current
(A)
I Ip(max)
I_Mon
Specification
100 50
1
54
0.03~0.07
63802
18
0.876~0.925
1
108
0.03~0.07
63803
36
1.771~1.829
1
135
0.03~0.07
63804
45
2.219~2.281
10.5.2 CC Mode Operation & Measurement Verification
1. Press CC to return to main screen of DC CC mode.
2. High current DC Source settings are shown as Table 10-9. Please follow the model under
CC mode to set loading, S.R. & I
p(max). According to the model to confirm if the values of
Meter & Load meet to that in Table 10-9.
Table 10-9
Source Setting Load Setting
Meter & Load
Measurement
Model
Voltage (V)
Current (A)
Current (A)
S.R.
Ip(max)
I Specification (A)
100 50
5
100
54
4.959~5.041
63802
18
17.946~18.054
5
108
4.923~5.077
63803
36
35.892~36.108
5
135
4.905~5.095
63804
45
44.865~45.135
10.5.3 CR Mode Operation Verification
1. Press CR to return to main screen of DC CR mode.
2. High current DC Source settings are shown as Table 10-10. Please follow the model
under CR mode to set R value & I
p(max). According to the model to confirm if the values of
Meter & Load meet to that in Table 10-10.