Remote Operation
Query Syntax : [LOAD:]CURR[:LEV][:AMPL][:DC]:FALL?
Response Parameter: <NR1>
Example : CURR:FALL 200
[LOAD:]CURR[:LEV][:AMPL][:DC]:RISE
Description: This command sets current rise slew rate of constant current mode.
Setting Syntax : [LOAD:]CURR[:LEV][:AMPL][:DC]:RISE
Setting Parameter : <NR1>, 63803/63804 Range: 4 ~ 600, 63802 Range: 4 ~ 240, Unit:
A/msec
Query Syntax : [LOAD:]CURR[:LEV][:AMPL][:DC]:RISE?
Response Parameter: <NR1>
Example : CURR:RISE 200
[LOAD:]CURR[:LEV][:AMPL]:HIGH
Description: This command sets the maximum RMS current. (AC CC mode only) (Don’t
support parallel mode)
Setting Syntax : [LOAD:]CURR[:LEV][:AMPL]:HIGH
Setting Parameter : <NR1>, 63804 Range: 0 ~ 45, 63803 Range: 0 ~ 36, 63802 Range:
0 ~ 18, Unit: A
Example : CURR:HIGH 45
[LOAD:]CURR:MAX[:LEV][:AMPL][:AC]
Description: I
rms
(max) setting, this command is only operating in AC CR Mode.
Setting Syntax : [LOAD:]CURR:MAX[:LEV][:AMPL][:AC]
Setting Parameter : <NR2>, 63804 Range: 0 ~ 45, 63803 Range: 0 ~ 36, 63802 Range:
0 ~ 18, Unit: A (AC CR Mode)
Query Syntax : [LOAD:]CURR:MAX[:LEV][:AMPL][:AC]?
Response Parameter: <NR2>
Example : CURR:MAX 45
[LOAD:]CURR:MAX[:LEV][:AMPL]:DC
Description: I
p
(max) setting, this command is only operating in DC Mode.
Setting Syntax : [LOAD:]CURR:MAX[:LEV][:AMPL]:DC
Setting Parameter : <NR2>, 63804 Range: 0 ~ 135, 63803 Range: 0 ~ 108, 63802 Range:
0 ~ 54, Unit: A
Query Syntax : [LOAD:]CURR:MAX[:LEV][:AMPL]:DC?
Response Parameter: <NR2>
Example : CURR:MAX:DC 45
[LOAD:]CURR:PEAK:MAX[:LEV][:AMPL][:AC]
Description: I
p
(max) setting, this command is only operating in AC CC/CP/RLC/RLC CP/RLC
Inrush Mode.
Setting Syntax : [LOAD:]CURR:PEAK:MAX[:LEV][:AMPL][:AC]
Setting Parameter : <NR2>, 63804 Range: 0 ~ 135, 63803 Range: 0 ~ 108, 63802 Range:
0 ~ 54, Unit: A (AC Mode)
<NR2>, 63804 Range: 0 ~ 200, 63803 Range: 0 ~ 160, 63802 Range:
0 ~ 80, Unit: A (Inrush Current Mode)
Query Syntax : [LOAD:]CURR:PEAK:MAX[:LEV][:AMPL][:AC]?
Response Parameter: <NR2>
Example : CURR:PEAK:MAX 135
[LOAD:]CURR:LIMit?
Description: This command queries the setting range of current. (AC CC mode only) (Don’t