Local Operation
After pressing Enter, the page of CP mode can be seen is shown as below. Users can press
the arrow key to select the setting parameters on the LCD display, including load power P
set
,
PF, I
p
(max), Time out and Cut off voltage. Users can type the values with the Entry keys.
V
=
0
. 0
0
T
=
0
0
: 0 0
: 0
0
: 0
0
0
m
s
The procedures of setting P
set
= 350.00, PF = 0.8, I
p
(max) = 35.0, TIMEOUT = 00:00:20 and
CUTOFF V = 21, are described as below:
1. Move the cursor to the command line of “P
set
=”.
2. Press 3 , 5 , 0 then press ENTER to change the value to “350.00”.
3. Move the cursor by pressing ( ◄, ► ) to the command line of “PF=”.
4. Press 0 , . , 8 then press ENTER to change the value to “0.800”.
5. Move the cursor by pressing ( ◄, ► ) to the command line of “I
p
(max)=”.