Performing a Test
118
Code chip
The code chip provides the meter with important information on the manufacturer-specific
parameters of the respective test strip lots. It contains information about the test method, the lot
number and the expiration date.
Each test strip box contains its own code
chip. Make sure you have the code chip at
hand before performing the first test with
a new test strip lot.
When you insert a test strip from a new lot
for the first time, the meter prompts you to
insert the corresponding code chip. At
this point, compare the code number you
see on the display with the number that is
printed on the test strip package you are
using. If the two code numbers are identi-
cal, insert the new code chip into the slot
in the meter.
Once inserted, the code chip information
is read and stored on the meter. The meter
can store up to 200 code chip datasets
(100 test strip lots and 100 quality control
lots). Further tests using the same strip lot
can be performed without inserting the
code chip again.
The code chip is no longer required, once
the data is stored. You may leave it in the
meter, or you can use it with other meters
using the same test strip lot.
Pat. Test - Ins. Chip
Insert Code
Chip
proBNPPar:
04/19/2016
2721

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)