Performing a Test
140
STAT tests
STAT tests are a limited number of tests that can be performed in emergency situations, see
page 110. If the meter is configured to perform STAT tests, you have the option of performing a
test by choosing STAT even though a QC Lockout is in place.
1 To perform the measurement without
performing a quality control test, touch
STAT, or:
2 Touch to perform the quality control
test before the measurement.
When a STAT test is performed, this infor-
mation is stored with the result. The
number of permissible STAT tests is
reduced by 1. After all pending quality
control tests are performed, the specified
number of STAT tests is available again in
case of a new lockout.
Pat. Test - Lockout
Test Parameter Locked!
Please Perform QC.
PID111SCHULZM457
SCHULZ, MANFRED
JONES, TOM
Pat:
Op:
proBNPPar:
STAT
PB1234Code:
04/19/2016
Pat. Test - Result
620
pg/mL
proBNP PB1234
SCHULZ, MANFRED
JONES, TOM
Par:
Op:
Code:
PID111SCHULZM457Pat:
STAT Test
04/19/2016
04/19/2016 10:17

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)