Meter Setup
104
Instrument quality control (IQC) lockout
An IQC Lockout also requires a quality control test with dedicated IQC test strips to be per-
formed. You may select from the following options:
No
Daily
Weekly
Monthly
All enabled IQC Lockouts, if the time period has elapsed, are activated at 8:00 a.m. (default set-
ting, time can be changed with a DMS).
See page 101 on how to access the Setup-QC Settings menu.
1 Touch IQC Lockout to set regularly recur-
ring instrument quality control tests (IQC
= Instrument Quality Control).
2 Touch the button to select the time interval
of choice. Your selection is now high-
lighted.
3 Touch to save this setting, or touch
to exit this menu without saving any
changes. The display automatically returns
to the previous screen.
QC Lockout
IQC Lockout
QC Result Format
Custom Range Trop. T
Reset Test Param.
Setup - QC Settings
04/19/2016
No
Daily
Weekly
Monthly
Select IQC Lockout
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)