Troubleshooting
182
Meter reset
A hard reset should only be performed if all other remedies have failed.
1 Place the meter on a level surface.
2 Press the button for at least
12 seconds.
The meter powers off and on again.
The Roche logo is displayed.
If the Roche logo does not appear within
60 seconds, place the meter in the Hand-
held Base Unit for a minimum of
15 minutes to recharge the battery pack.
The meter performs a system check.
The screens for entering the date and the
time appear.
3 Enter the date and time. After you have
entered the correct information, confirm
each screen with .
Unless all lockouts are disabled (see
chapter 3.4), the meter will now be in QC
Lockout due to the manually entered
date/time.
4 Synchronize the meter’s date/time with
the date/time of your facility via the Hand-
held Base Unit or, if working wirelessly,
wait at least ten minutes for the next
WLAN synchronization before performing
any further tests.
Even if your configuration does not require it,
we recommend always performing a QC test
after a meter reset.
Self-Check
Performing self-checks...
04/19/2016
SW Version: v04.00.00
09:15 am
Self-Check
Self-checks complete.
04/19/2016
SW Version: v04.00.00
09:15 am
Set Date
Month:
1
Day:
1
Year:
2016
04/19/2016
Minutes:
49
Hour:
3
Set Time
04/19/2016
AM PM

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)