Meter Setup
106
Custom Range Troponin T
The quality control range for Troponin T can be customized to comply with local guidelines. The
Custom Range Trop. T function enables you to narrow the default range.
Default Range: The meter displays the quality control range provided by Roche in the
code chip.
Custom Range: The option Custom Range lets the user define their own quality control
range within the default range.
See page 101 on how to access the Setup-QC Settings menu.
Do not use the Custom Range option when using level 1 controls. The target value is outside
of the quantitative measuring range.
1 Touch Custom Range Trop. T to set the
quality control range.
You may select from the following options:
Default Range (Range provided in the
code chip is selected and displayed.)
Custom Range (The allowed percentage
deviation from target value can now be
customized).
2 Touch the button with the setting of
choice. Your selection is now highlighted.
If you selected Default Range, touch
to save this setting, or touch to exit this
menu without saving any changes.
If you selected Custom Range, touch
to proceed with corresponding settings, or
touch to exit this menu without saving
any changes The display automatically
returns to the previous screen.
QC Lockout
IQC Lockout
QC Result Format
Custom Range Trop. T
Reset Test Param.
Setup - QC Settings
04/19/2016
Custom Range
Default Range
Custom Range Trop. T
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)