Monitor Troubleshooting
The following suggestions may be useful in troubleshooting monitor
malfunctions:
Problem
Check
No Image
Is the power cord is properly
connected?
Is the video cable is properly
connected?
Are any pins on the video cable
bent?
Is the video signal present on
the backplane?
Warning message: “Wrong
monitor type or monitor
firmware detected! ...” appears
after startup.
This referenced non-existant
pages in the manul. Check the
online help?
Picture is fuzzy.
Adjust the picture contrast and
picture brightness. Some SVGA
cards having an excessive video
output level will cause a fuzzy
picture at the maximum
contrast level.
Video test patterns are not
clear, bright, parallel or square.
Replace the monitor.
Load Default Monitor Settings
To load the default monitor settings:
1. Press the Utilities key on the control console.
2. In the “Utilities” menu touch the Monitor button to display the
"Monitor Menu".