GE Voluson E10 Training Manual
© 2017 Conquest Imaging
60
Touch Panel Not Responding
Calibrate the Touch Panel.
Reload software.
Replace Upper Panel Assembly.
Probe Recognition
Check all the probes on all the ports. Remove all the probes, and then
check each probe singly in every port.
If the problems persist with all the probes,
If only one probe fails to be recognized, replace the probe.