GE Voluson E10 Training Manual
© 2017 Conquest Imaging
8. To exit the Monitor Test program, press ESC.
Trouble Shooting Operator Panel Issues
This section covers common operator panel issues.
No Audio
Check volume settings in the application and also in Windows.
Use headphones, (the type that you use on a personal CD player or a
laptop) to test the audio output directly from the back of the BEP.
If no Audio is present on the BEP, reload software. If the problem
persists, replace the BEP.
Always measure the speakers’ impedance; it should measure
approximately 7 ohms. If speakers are bad, it is possible that the
amplifier on the on the UI may be defective, too.
No Video on LCD Display
If the video is too dim, has dimmed areas or there is no video at
all, replace the HV LCD inverter.
Check the cabling within the system.
If the HV LCD inverter is burnt, there is a possibility that the
Digital Video Card on the PC also got damaged; in that case you
can replace the Video Card.
Replace the Video Adapter inside the BEP.
Replace the Back End Processor if the problem continues.
Replace the Upper Panel Assembly.
Wrong Key Activated on the Touch Panel
Calibrate touch panel.
Replace Upper Panel Assembly.