Spider DSA User’s Manual
144
The following two methods can be used by the user if they want to add the
customized markers to all the signals present on the display window.
o Place a vertical cursor and click the iconAdd marker at cursor” on
the hidden toolbar of the display window.
o Right-click on the display window and select Add
Cursor/MarkerCustomized markers.
Saving and Recording Data
As previously described, there are two methods of data retention during a test:
Saving and Recording. A save is activated by pressing the Save button on the
control panel, which saves block signal data and figures of display windows to the
local disk. Press the Rec. button to activate recording, which records time stream
data to the internal flash memory.
Although EDM supports up to 24 Spider modules with 192 input channels,
recording to flash memory has certain limitation when highest sampling rate of
102.4 kHz is selected. At the sampling rate of 102.4 kHz, the recording to flash
memory is limited up to 144 channels (6 channels * 24 modules). At sampling rate
of 81.92 kHz or lower, the recording to flash memory can apply to 192 channels at
maximum.