Spider DSA User’s Manual
349
Reset All Alarms button is to reset all the alarm counting. Alarm counter will
be reset to zero.
Setup Trigger gives the option of setting up a trigger-based action with the help
of event action rules. Only the input channels on the master module can be set as
the trigger source. After the trigger setup is initially finished, the trigger switch
will be turn on automatically.
Preview Trigger button is to pop up the Waiting for Trigger window in case you
hide it previously.
Data Used (MB) is the total size of data in megabytes for all tests in the project.
Trigger Switch toggle button is to switch ON/OFF trigger.
Start All: start running all the tests.
Stop All: stop running all the tests.
Download from All button is to download all data files from all Spider devices.
Clear All Alarm Status button is to reset the alarm status shown on the third
tab of the control panel.
Output Setup
The second tab on the control panel is for output channel setup. Here you can
configure one or more output channels as shaker drive or other DUT stimulation
signals. Since closed loop control is not available, the user should be careful to
stay within the safety limits of the shaker/amplifier and test object when
configuring the output levels. All enabled channels will send out drive signals
while the test is running.