Spider DSA User’s Manual
174
Select Black Box Setup from the Setup menu and press Refresh.The test will be
listed under Uploaded Tests.
Figure 121: Black Box Setup Dialog box
When the check boxAlways overwrite first entry when running a test” is checked,
the active running test will always be automatically uploaded to the front-end as
one of the tests that can be run in the Black Box mode. If this option is not
checked, the Black Box table will not be updated.
By default, the test in position 1 will run. This can be changed from the front panel
LCD display.
Tests stored in the front-end box can be executed by several means: an iPad
connected to the box through wireless, via the digital I/O, or via the front LCD
panel.
Run the Test with Front Panel
To select which test to run, press the right arrow key on the Spider’s front panel
until the Test List is shown. Then use the up and down arrow keys to select a test.
When the Start button is pressed, the selected test will be run.