To use this feature, simply check the fifth or sixth option as shown below to
activate the exceeded limit recording. This setting can be found under the Test
Configuration->Save/Recording Setup menu, and click the Record Time Stream
tab.
The limit can be bound to any channel on the Spider master device. When the
limit exceeded event (spectrum or block) is set to start recording, the same event
cannot be used to stop recording. So, the recording can be stopped by digital
input, stop recording button, or stop test button.
When the record duration is set to zero, the recording will not stop unless the user
or system action stops it. When the record duration is set to non-zero value, the