HYBRID Coater Service Manual ver1.1 Chapter 5
45
23 Machine calibration :
Refer to User Manual of the Hybrid C:
You can choose between a full calibration and a limited calibration (check the box!).
A full calibration is needed
When something has been replaced or
On request of the customer
A limited calibration, this can be:
Mapping the motion system
Calibrating the feeder positions when needed
One pipette
NOTE: by calibrating one part of the HYBRID it may be necessary to calibrate other parts
which are dependant of this part.
24 Check positions of units/ syringes :
Check if the positions of the units/ syringes are accurate via the software.
25 Perform an overall function test !